Quick Start Guide
How to use the Stryker Multigen Radiofrequency Generator Quick Start Guide
Step by Step
Stryker Multigen Basic Setup Components
- Stryker Multigen Pain Management RF Generator
- Power Cable
- Connecting Cable(s)
- Ground Pad Cable
- Ground Pad (disposable)
- Monopolar or SGS Electrode (sterilization required)
Blue = 50 mm length, Red = 100 mm length, Yellow = 150 mm length - Cannula(s) (single-use, disposable component)
NOTE: The color of the HUB identifies the length of the cannula while the color
of the STYLET identifies the gauge of the cannula.
Blue = 50 mm length, Red = 100 mm length, Yellow = 150 mm length
White = 18 gauge, Yellow = 20 gauge, Black = 22 gauge
Stryker Multigen Radiofrequency Procedure Protocol:
- Turn on the Multigen generator (press top-left button). The unit will perform a self-test automatically.
- Choose SELECT SAVED PROCEDURE or USE DEFAULT SETTINGS.
- From the SELECT SAVED PROCEDURE screen, select the correct folder and file to initiate the RF procedure.
- After selecting the desired file, the RFPROCEDURE SUMMARY SCREEN will appear. Confirm the TIME and TEMPERATURE settings are correct, then press ACCEPT/BEGIN. See IFU for more details.
- Attach the Stryker RF generator GROUNDING PAD CABLE to the corresponding port on the RF generator.
NOTE: Align RED dot to RED dot to connect the cable properly.
- Place the RF GENERATOR GROUNDING PAD on the patient at the physician`s direction. Connect the pad to the GROUNDING PAD CABLE.
- Attach the CONNECTING CABLE(S) to the corresponding port(s) on the generator.
- At the physician`s direction, grasp the sterile end of the appropriate RF ELECTRODE and attach it into the corresponding CONNECTING CABLE.
NOTE: If using the ID TAG SYSTEM, attached the numbered RF ELECTRODE to the corresponding numbered port (1-1, 2-2, 3-3, 4-4). See Instructions for use for more details.
- The pain management physician will now place the RF ELECTRODE into the designated CANNULA.
NOTE: The IMPEDANCE value will now be displayed on the Multigen generator.
- Stryker MultiGen RF Generator ONLY: Select the appropriate RF CANNULA and RF ELECTRODE placement on the screen by touching the corresponding TAB on the RF generator console. See Instructions for use for more details.
SENSORY STIMULATION: SENSORY stimulation may be performed to confirm the cannula is properly placed along the painful (target) nerve. The Sensory Stim screen is Green and the corresponding button located on the far right of the screen is labeled SENSORY. Press START when ready to stimulate. Press the ARROWS on the screen to adjust the voltage. Press STOP when testing is complete.
MOTOR STIMULATION: MOTOR stimulation may be performed to confirm the cannula is not placed on a motor nerve. This Motor Stim screen is Orange and the corresponding button located on the far right of the screen is labeled MOTOR. Press START when ready to stimulate. Press the ARROWS on the screen to adjust the voltage. Press STOP when testing is complete.
LESION MODE: When the physician is ready, press the Blue LESION button to access the RF OUTPUTscreen. Verify TEMPERATURE and TIME settings are correct prior to pressing START. If required, EDIT the settings of each placement by pressing the EDIT button. To begin RF lesioning press the START button. To guard against any accidental starts, there will be a second pop-up window where you will have to confirm START. Once the TEMPERATURE is within two degrees of the “set temperature”, the TIME will begin to count down. The lesion is complete when the TIME runs out.
WARNING: Only trained and qualified pain management physicians should use a Stryker Multigen pain management radiofrequency generator. Before using the Stryker RFA Machine or any component of the RFA generator system, read and understand the instructions for use supplied with the RF equipment. Please call 858-336-1377 if you have any questions.